Ranger Irrigation Fluid Warming System

Model 247 Product Safety Features

The following chart describes the safety alarm features of the Ranger Irrigation warming unit.
ALARM TyPE
ALARM TyPE
Primary over-temperature
Primary over-temperature
alarm system
alarm system
Independent back-up
Independent back-up
over-temperature alarm
over-temperature alarm
system (cut-off).
system (cut-off).
Under-temperature alarm
Under-temperature alarm
:
The difference in over-temperature alarm system cut-off points does not indicate any difference
note
in safety and is simply a result of tighter controller tolerances.
The Ranger Irrigation Fluid Warming System is designed with very tight controls to regulate its temperature
to the 41°C set-point. The microprocessor controller measures the temperature at the fluid pathway inside the
warming plate four times every second, making it very responsive to changes in flow rate.

SET-POINT
SET-POINT
INDICATION
INDICATION
48°C
48°C
LED alternately displays
LED alternately displays
"HI" and temperature,
"HI" and temperature,
alarm sounds, alarm light
alarm sounds, alarm light
flashes.
flashes.
50°C
50°C
LED displays the
LED displays the
temperature or has gone
temperature or has gone
dark, alarm sounds.
dark, alarm sounds.
33°C
33°C
LED displays "LO"
LED displays "LO"
and temperature, alarm
and temperature, alarm
sounds, alarm light
sounds, alarm light
flashes.
flashes.
DESCRIPTION
DESCRIPTION
Primary Alarm: Alarm
Primary Alarm: Alarm
sounds, controller cuts
sounds, controller cuts
power to the heaters until
power to the heaters until
temperature falls below 48°C.
temperature falls below 48°C.
When heater temperature
When heater temperature
drops below 48°C, audible
drops below 48°C, audible
and visual alarms stop.
and visual alarms stop.
Primary failure. Temperature
Primary failure. Temperature
will rise to 50°C within a few
will rise to 50°C within a few
seconds. The safety backup
seconds. The safety backup
system is activated at 50°C
system is activated at 50°C
and power is immediately shut
and power is immediately shut
off to the
off to the
heating plates.
heating plates.
Heater temperature has fallen
Heater temperature has fallen
to 33°C. Alarms will stop
to 33°C. Alarms will stop
when temperature rises above
when temperature rises above
33°C. Continue use of unit.
33°C. Continue use of unit.