Producer's guide and declaration - Electromagnetic IMMUNITY
The R26M/R36M is suitable for use in the specified electromagnetic environment. The purchaser
or user of the The R26M/R36M should assure that it is used in an electromagnetic environment as
described below:
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Radiated
electromagnetic
field
IEC 61000-4-3
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
50
IEC 60601-1-2
Compliance
Test level
8 kV contact
IEC 60601-1-2
2/4/8/15 kV air
3 V/m
IEC 60601-1-2
80 MHz to 2.7 GHz
2 kV for power
IEC 60601-1-2
supply lines
1 kV for input/output
lines > 3 m
0.5/1 kV differential
IEC 60601-1-2
mode
0.5/1/2 kV common
mode
3 V
IEC 60601-1-2
150 kHz to 80 MHz
Electromagnetic
Level
Environment
Floors should be wood. con-
crete or ceramic tile. If floors
Test level
are covered with synthetic
material. the relative humidity
should be at least 30%.
Portable
Test level
communications
should be used no closer to
any part of the R26M/R36M
including cables.
Recommended istance:
d = 1,2√P 80 MHz to 800 MHz
d = 2,3√P 800 MHz to 2,5 GHz
Where P is the aximum power
output of the transmitter
in watt (W) based on the
transmitter's producer and d is
the recommended distance in
meters (m).
Mains power quality should be
that of a typical commercial or
Test level
hospital environment.
Mains power quality should be
that of a typical commercial or
Test level
hospital environment.
Portable and mobile RF com-
Test level
munications
should be used no closer to
any part of the R26M/ R36M
including cables.
Recommended distance d =
1,2√P
and
mobile
RF
equipment
equipment