be strictly performed only by qualified technical personnel, duly authorized by the manufacturer and
implementing the appropriate apparatus for checking the levels and frequencies of all the output
signals by the transducers supplied on-issue with the device. Once device calibration testing using
the special purpose apparatus is complete, it is necessary to run a device function test.
Device calibration is warranted by the manufacturer only if the transducers implemented are directly
connected to the device.
In the event of replacement of transducers and/or other device components, exclusive use of
Resonance Original spares supplied directly by the manufacturer is highly recommended. Whenever
transducers are replaced, recalibration of the device by qualified technical personnel is necessary.
The device is provided with a lithium battery type CR2032 which enables constant updating of the
date and time data displayed by the device. Said battery is welded onto the device PCB and possible
replacement thereof must strictly be performed only by qualified technical personnel. Failure to
observe this instruction may result in damages during battery replacement, thus preventing correct
display of the date and time data by the device.
Blown fuses must be replaced only with the same type of fuses, as specified on the device itself.
Maintenance and cleaning schedule
Daily
Presence of signals in the transducers
provided
Correct probe signals operation using the
calibration cavity provided
Probe cleanness*
Monthly
General device condition
O-Ring probe status
*After every cleaning procedure performed on probe, the Compliance Calibration is mandatory.

Environmental conditions

Storage and handling
Operating conditions
Weekly
Annually
temperature range
from -20 °C to +50 °C / from -4 °F to +122 °F
humidity range
from 20% to 80%, without condensation
pressure range
from 500 hPa to 1060 hPa
temperature range
from +15 °C to +35 °C / from +59 °F to +95 °F
humidity range
from 30% to 90%, without condensation
pressure range
from 700 hPa to 1060 hPa
Good condition of all accessories cables and
related jack-plugs
Compliance calibration*
Perform calibration and routine
maintenance check
45