Resonance R26M Série Mode D'emploi page 42

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

NOT FOUND
It appears during middle ear
analyzer reflex testing, when
the automatic peak detection
function has been set but the
peak is not detected
LEAKAGE TEST
It appears during middle ear
analyzer testing at the set start
pressure, prior to plotting of the
tympanometric graph
LEAKAGE
It appears during middle ear
analyzer tests when a pressure
leakage is detected
NO REFLEX
It appears during threshold tone
decay testing, when a prior, valid
stapedial reflex performed pre-
viously at the same frequency,
is missing
TEST NOT
It appears during middle ear
COMPLETED
analyses when the tympano-
metric graph has not been
completed correctly
POSITIVE
This message appears during
PRESSURE
performance of the ETF testing
for non-perforated eardrums,
whilst positive pressure is being
generated
NEGATIVE
This message appears during
PRESSURE
performance of the ETF testing
for non-perforated eardrums,
whilst negative pressure is being
generated
42
Error
It means that during
message
plotting of the tympa-
nometric graph, the
compliance variation
levels were not enough
to allow the device
to detect the point in
which to perform the
reflex test
System
It means that the
message
device is running an
automatic pressure
maintenance check
prior to performing
the test.
Error
It means there is a
message
pressure leak in the
device's pressure circuit
Error
It means that no
message
valid reflex was found
during previous reflex
testing, at the same
frequency set for the
decay test
Error
It means that the
message
tympanometric graph
has not been plotted
correctly and the test
has therefore been
interrupted
System
It means that positive
message
pressure is being
generated for test
performance
System
It means that negative
message
pressure is being
generated for test
performance
Check to ensure that
pressure is being
generated by the device;
check that the probe exit
is not obstructed; check
to ensure that the probe
has been inserted into
the ear and not into a
rigid cavity; the message
may occur due to a
pathology by the patient
being tested: in case of
doubt, please repeat the
test on a norm-auditory
subject.
No action is required
Check to ensure that
the probe has been well
inserted into the test
cavity; ensure that the
cone-shaped section
of the actual probe has
been firmly closed; check
that the dimension of the
tip being used is appro-
priate for the cavity the
probe has been inserted
into.
If this is not due to
pathology by the patient
being tested, it is neces-
sary to perform a reflex
test for the frequency
levels and stimulation
side that the decay test
must be performed on
Check to ensure that
there aren't any pressure
system leaks
No action is required
No action is required

Publicité

Table des Matières
loading

Table des Matières