Technical Information
Procedure to Connect the Serial and I/O ports
to an External Cable

Window-Meanings

Table 6
N.
000
008
100
101
102
103
104
105
106
107
108
288/306
Window-Meanings
Read/Write
Datatype
R/W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
R/W
N
R/W
L
R/W
L
R/W
N
Turbo-V 903 Pumping System User Manual / 87-901-016-01
Description
Start/Stop (in remote mode
the window is a read only)
Remote (default)
or Serial configuration
Soft Start (write only in Stop
condition)
Set Point type
Set Point threshold
(expressed in Hz, mA or s)
Set Point delay: time
between the pump start and
the set point check (seconds)
Set Point signal activation
type: the signal can be "high
level active" or "low level
active"
Set point hysteresis
(in % of threshold)
Water cooling
Active Stop
(write only in stop)
Baud rate
Admitted Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 867)
0 to 99999
(default = 0)
0 = high level
active
1 = low level
active
(default = 0)
0 to 100
(default = 2)
0 = NO
1 = YES
0 = NO
1 = YES
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
(A)