Battery replacement

Check the batteries charge regularly. Low battery voltage (de-
pending on consumption) leads to light intensity decrease and
possible appearance of LED lamps flashing. In both cases, it's
necessary to replace the batteries. When replacing, use only
new, high quality alkaline batteries (or fully charged batteries).
Besides, observe manuals' requirements for use of
the batteries.
Installation of
Installation of
batteries for
batteries for
KaWe EURO-
KaWe EURO-
LIGHT, 2 pcs.
LIGHT
medium size
1.5V (type C)

Lamp replacement

The correct performance of the ophthalmoscope with original
light quality is only ensured when using original KaWe replace-
ment lamps.
KaWe EUROLIGHT E10
1. Unscrew the retaining screw on the handle and remove the
ophthalmoscope head from the handle.
2. Unscrew the cap ring (1).
3. Pull out the lamp holder (2) on the bottom of the head and
replace the lamp.
4. Clean the glass bulb (3) of the new lamp with alcohol. The
glass bulb must be clean and free from fingerprints (grease
free).
5. Install the lamp holder and lamp back into the head.
Installation
6. Screw the cap ring.
of batteries
KaWe EUROLIGHT E15/ E16/ E25/ E30/ E35/ E36
for KaWe
1. Remove the ophthalmoscope head from the handle.
PICCOLIGHT, 2
1.1. Models E15/E16:
pcs. minion type
Unscrew the retaining screw on the handle.
1.2. Models E25/E30/E35/E36:
1.5V (type AA)
Push down on the wickered sleeve.
2. Remove the lamp at the bottom of the head.
3. Clean the glass bulb of the new lamp with alcohol. The glass
bulb must be clean and free from fingerprints (grease free).
4. Insert the lamp into the head until it stops.
KaWe PICCOLIGHT E50/ E56
1. Unscrew the ophthalmoscope head from the handle.
2. Remove the lamp at the bottom of the head.
3. Clean the glass bulb of the new lamp with alcohol.
The glass bulb must be clean and free from fingerprints
(grease free).
4. Insert the lamp into the head until it stops.

9. Accessories and Spare parts

Accessories and spare parts can be found in the general catalog
of our company at https://www.kawemed.com/download/
broschueren/ or can also be viewed at your specialist dealer.

10. Disposal

The product is to be disposed in accordance with local environ-
mental regulations. Disposal instructions are also shown on the
pictograms on the device or packaging.
Defective and/or recyclable electrical / electronic
devices and rechargeable batteries must be trans-
ferred to a designated waste collection point for
recycling.