KaWe PICCOLIGHT E50/ E56
1
2
(1) Light outlet
(dustproof)
(2) Blend wheel
(except E50 model)
(3) Inspection window
(4) Correction lens wheel
with 18 lenses (± 20 diopters)
a) Install batteries into the handle (2 pcs. AA 1.5V minion).
Observe the polarity. For detailed information see Section 8:

Maintenance and Service

b) Screw the head onto the handle (6).
c) Move the slide switch (7) to turn the lamp on and off .
d) Battery and lamp replacement is possible after unscrewing
the head. For detailed information see Section 8:
Maintenance and Service

5. Preparing for operation

• Periodically check that all parts are in good condition -
replace defective parts.
• Check the smooth running of moving parts (e.g. correction
lens /blend wheel, plug connector).
• Keep all electrical contacts clean.
• Check the batteries charge.
After a successful inspection, the ophthalmoscope is ready for
use.

6. Hygienic preparation

Keep the ophthalmoscope in its packaging to protect it from dirt
and dust.
16
3
4
7
5
6
(5) Correction value indication
(+) black; (-) red
(6) Thread
(7) On/off slide switch

Manual cleaning/disinfection

Protect the device from moisture!
Do not place the ophthalmoscope in liquids, make
sure that no liquid gets into the casing.
Clean the outside of the ophthalmoscope with a damp, soft,
lint-free cloth. If necessary, wipe the light outlet glass with a
cotton swab slightly moistened with alcohol.
• Do not press on the glass.
• Verwenden Sie keine scheuernden Reinigungsmittel!
Do not use abrasive cleaning agents. For disinfection, the opht-
halmoscope can be wiped with an alcohol-based surface disin-
fectant. Use a slightly damp cloth.

Automatic cleaning

Not applied.

Sterilization

Not applied.

7. Transportation and storage

Ophthalmoscopes are to be stored and transported protected
from dust, moisture and contamination.

Environment parameters

Temperature
+10°C to
Operation
+35°C
Storage
-10°C bis
+55°C
Trans-
portation
8. Maintenance and Service
The end of the useful life is usually determined by wear and tear.
Damaged parts are to be sorted out and replaced immediately.
If the light intensity decreases, check for the following possible
causes:
1. Battery charge; replace or recharge if necessary.
2. Clean or replace the lamp.
If the above reasons are excluded, repair or replacement may be
required. In this case, please, contact a specialized dealer.
Relative
Air
humidity
pressure
800 GPa to
30% to 90%
1060 GPa
700 GPa to
1060 GPa
10% to 95%
500 GPa to
1060 GPa