Do not use solvents such as acetone, alcohol or similar for cleaning.
Use only the cleaning agents and disinfectants listed below.
Excessive dosing and extended exposure times that deviate from
the specifications described here or from the specifications of the
manufacturer of the cleaning and disinfection agent, can cause
damage to the breathing connection, e.g. the visor can become
brittle.
Do not use a cloth to clean or dry the inside of breathing
connections with visors that are marked "PC/CC", as the coating
can be damaged as a result. After drying, these breathing
connections should be aired unpacked for at least one day under
standard climatic conditions according to DIN EN ISO 291, so that
the coating can regenerate and the optimum working life is
achieved.
Clean and disinfect the breathing connection:
Clean the breathing connection using water (max. 30 °C) with a
universal cleaning agent, Sekusept Cleaner
solution). Immediately afterwards, remove the cleaning agent
completely under running water.
Immerse the breathing connection for 15 minutes in the
disinfectant bath (max. 30 °C) with Incidur
(2 % solution). Immediately afterwards, remove the disinfectant
completely under running water.
Leave the breathing connection to dry thoroughly in a drying
cabinet or in the open air. Temperature max. 60 °C. Do not expose
to direct sunlight.
Do not lay the mask / helmet adapter on the face seal.
Grease the guides of the mask / helmet adapter with Vaseline.
14
CAUTION
®1)
(0.5-1.0 %
®1)
disinfectant
Testing
Test the breathing connection in accordance with the test and
maintenance intervals.
Test and maintenance intervals
The following details are recommendations based on the BGR 190
applicable in Germany. Observe your own national guidelines.
Work required on a full face
mask
Check by the wearer
Visually inspect, check correct
function and leak-proofing
Cleaning and disinfection
Replace the exhalation valve
disc
Replace the speech diaphragm
1)
For airtight packed masks, every 2 years.
2)
For airtight packed masks, otherwise every 6 months.
Visually inspect the speech diaphragm
Remove the speech diaphragm (see "Replacing the speech
diaphragm or O-ring" on page 16).
Visually inspect the inhalation valve
The inhalation valve is visible in the connector through the visor.
Check by blowing into it that both wings of the valve disc move
freely. If necessary, replace the valve disc.
X
X
1)
X
X
2)
X
X
X