Technical data
Protection against solid
objects:
Clinical test:
Maximum deviation of cuff ± 3 mmHg or 2% of the reading
pressure measurement:
Maximum deviation of
pulse rate display:
Applicable standard:
IP20
IP classification is the degree of protection provided by
enclosures in accordance with IEC 60529. This device is
protected against solid foreign objects of 12mm
diameter and greater such as fingers. This device is not
protected against water.
Accuracy complies with the requirements
of ISO 81060 Part 2
(whichever is the higher)
± 5 %
IEC 80601-2-30 : "Particular requirements for basic
safety and essential performance of automated
non-invasive sphygmomanometers"