Safety
 Before using Traction Frame, it is essential to have a thorough understanding of this manual. This manual contains instructions for general use and
guarantees improved safety.
 The KINETEC
system is according to NF EN 60601-2-52 –June 2010. Directive 93/42/CEE from June 14 1993.
®
 The Kinetec system must not be used with traction frame components from another company.
 Before any use, please make sure that the weight of the patient and of the accessories are compatible with the security load of the bed.
 Check before each use and each week the fixing of the jaws and the general state of the accessories.
 The physician must define the protocol and ensure that it is correctly implemented.
 The weights must be positioned outside of the line of the frame ensuring that they do not pose a risk to any person interfacing with the frame.
 Do not position weights above the patient.
 When the traction frame is in use ensure that frame functions are locked out. All frame functions can be used when the traction frame is no longer in
use.
 Remove the traction frame when it is not used.
 The side barriers of the bed must be in high position when the frame is used.
 Failure to do so could result in patient injury or material damage.
Assembly
 How to use the Clamp EM and EF:
Position the EF on the BC120 at 5cm (2") from the end of the bar.
Position the EM on the same bar; make sure that the part (b) of the EM
is on the same line as the part (a) of the EF as to create a vice.
Make sure that the screw (c) is unscrewed but NOT TAKEN OFF.
Position the BC120 with the EM and the EF on the foot or the head of
the bed:
-
First, position the EF below the lower bed bar.
-
Then slide the EM above the upper bar and tighten the button (d).
-
Tighten the screw (c).
To ensure that the stability stays good enough, please check the
tightening of the screw (c) each week.
 Please respect the set up of the frame components, as indicated opposite, make
sure that the DME Equi are positioned on the top. The U bar must be positioned
above the BC120. The BU95.5 must be positioned must be positioned so that the
bars of U are outside the barriers of the bed.
Using
 During handling of the barriers and movements of the mattress, to take all the precautions in
order to avoid the pinching and wedging.
 During the using or during the moving of the bed with the frame, make sure that no
components are outside of bed frame (bars, accessories).
 Make sure that during the moving of the bed or during the handling no outer components can
knock against the frame (lighting, TV...).
 Do not move the bed with the frame and the patient, but in an emergency situation take care
of the obstacles which can provoke weight vibrations (elevator entrance).
 When you move the bed with the frame and the patient, please take care of for example the
elevator entrance.
Cleaning
 Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic and metal
components). We advise to pay particular attention to the articulations and on holding back zones like the screws and bolts.
 In order to ensure optimal hygiene, you are advised to clean the pads and the frame for each new patient. All the consumables enable hazard-free
disposal. We advise also to change the yellow balls (SP01) with each different patient.
 Recommendations to obtain a maximum hygiene of the pads.
Sterilisation of the pads ( if necessary) : Sterilised at 134 °C for 18 minutes.
Disinfecting of the pads: Washing at 30°C with use of a disaffection solution during the rinsing cycle. Example of product which can be used: Solution
"Baclinge" at 0.125% or "Souplanios" at 0,125% from ANIOS Laboratory. A complete list of distributors in your country is available on request.
Technical specifications
 Product
- Medical terms class 1
- Anodising aluminium tubular and epoxy aluminium jaw
 Environment
- Storage/transport conditions:
- Operating conditions:
Warranty
The KINETEC warranty is strictly limited to the replacement free of charge or repair in the plant of the component or components found to be defective.
KINETEC guarantees its Traction frames systems for 2 years against all defects of manufacture from the date of purchase by the consumer.
KINETEC is the only organisation able to assess the application of the warranty to its systems.
The warranty will be considered null and void if the device has been used abnormally or under conditions of use other than those indicated in the user's
manual.
The warranty will also be considered null and void in the event of deterioration or an accident due to negligence, inappropriate surveillance or
inappropriate maintenance, or due to transformation of the equipment or an attempt to repair the equipment.
ENGLISH
Before use, please read this document.
Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
Room temperature: 10 to 40°C / 50 to 105°F
Relative humidity: up to 80%