Indexing And Matching Sample Vials; Indexing A Single Sample Vials - Lovibond TurbiCheck Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Indexing and matching sample vials

GB
Matched sample vials are required to minimise the effects of optical variation from glass vial to
glass vial. Alternatively an indexed single sample vial can be used for every measurement.
Once vial orientation in the sample chamber is located, always use this new orientation mark
for all measurements independent of the white pre-printed triangle mark.
Indexing a single sample vial
Note:
Assume the indication "Underrange" is equal to 0.00 NTU.
38
1.
Fill a clean and dry vial up to the mark with dilution
water (see chapter: Preparing Dilution Water).
2.
Cap the vial.
3.
Hold the vial by the cap and clean it with the supplied
cleaning tissue to remove water spots and finger
prints.
4.
Switch the turbidimeter on.
5.
Place the vial in the sample chamber.
6.
Make sure that the marks are aligned.
7.
Place the cover on the instrument.
8.
Press [Read] key.
9.
Note the displayed result.
10.
Rotate the vial in the sample chamber about 45°.
11.
Place the cover on the instrument.
12.
Press [Read] key.
13.
Note the displayed result.
14.
Repeat this procedure until the smallest NTU value is
obtained.
15.
Mark the vial at this point. Use this orientation mark
for all further measurements.
TurbiCheck_1b 01/2011

Publicité

Table des Matières
loading

Table des Matières