Cod Vario; Preparing The Sample; Measurement - Lovibond MD100 CSB Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Methods
GB

COD vario

Select the appropriate vial for the desired range:
LR-Range:
0 –
MR-Range:
0 – 1500 mg/l, Order code 2420721
HR-Range:
0 – 15 000 mg/l, Order code 2420722

Preparing the Sample

Open a reaction vial with a white cap and add the specified volume.
(Ensure that appropriate personnel safety equipment is used.)
LR-/MR:
2 ml water sample
HR:
0.2 ml water sample
Prepare a blank (Note 1) by using deionised water (TOC-free) instead of the sample
(LR/MR: 2 ml, HR: 0.2 ml).
Replace the cap tightly. Invert the vial gently several times to mix the contents (The vial
will become hot during mixing!) and digest the vials for 120 minutes in the reactor at
a temperature of 150 °C. Remove the vials from the reactor and allow them to cool down
to 60°C or less. Mix the contents by inverting each vial several times while still warm. Then
allow the vials to cool to ambient temperature before measuring.
On
Off
RANGE
Mode
RANGE
Zero
Test
RA NGE
18
150 mg/l, Order code 2420720

Measurement

Fix the adapter for 16 mm vials on the sample chamber.
Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key.
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test range, the instru-
ment memorizes the last range used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test range first.
The display shows the following:
Place the blank in the adapter (Note 1 – 4) making sure that the
marks
are aligned.
Blanks are specially prepared for each individual test range.
Press the [ZERO/TEST] key.
The "Range" symbol flashes for approx. 8 seconds.
MD100_2b 05/2011