Lovibond MD100 CSB Mode D'emploi page 22

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Calibration Mode
GB
RESULT
CAL
Mode
Zero
Test
CAL
RESULT + x
On
Off
:
:
Note: The instrument calibration in the range MR is automatically taken also for HR.
22
The result is shown in the display, alternating with CAL.
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
By pressing the [ON/OFF] key, the new correction factor is calculated
and stored in the user calibration software.
Confirmation of calibration (3 seconds).
MD100_2b 05/2011

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Md100 codMd100 dcoMd100 doc

Table des Matières