Setting Date And Time; Calibration Mode; User Calibration - Lovibond MD100 CSB Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Menu options – Calibration Mode
GB
E 132
Store
Date
2
3
Time
Cal
Mode
SET
DATE
YYYY
(2 sec.)
Mode
Zero
Test
!
Store
Date
Cal
4
Time
Cal
cAL
CAL
CAL
LR
Zero
Test
RANGE
0.0.0
CAL
Zero
Test
R ANGE
MD100_2b 05/2011
If the instrument is not able to communicate with the IRiM, a timeout
occurs after approx. 2 minutes. The error E 132 is displayed for approx.
4 seconds. Subsequently, the instrument switches to test mode (see
also IRiM manual).
3
Setting date and time (24-hour-format)
2
After confirming the selection with the [MODE] key the value to be
edited will be shown for 2 sec.
The setting starts with the year (YYYY) followed by the actual value
to be edited. The same applies for month (mm), day (dd), hour (hh)
and minutes (mm). Set the minutes first in steps of 10, press the [!]
key to continue setting the minutes in steps of 1.
Increase the value by pressing the [MODE] key.
Decrease the value by pressing [ZERO/TEST] key.
Proceed to the next value to be edited by pressing [!] key.
After setting the minutes and pressing the [!] key the display will show
"IS SET" and the instrument returns to the measurement mode.

User calibration

4
Note:
user calibration (Display in calibration mode)
factory calibration (Display in calibration mode)
After confirming the selection with the [MODE] key the instrument will
show CAL/LR.
Scroll through ranges using the [MODE] key.
Place the blank in the adapter making sure that the marks
aligned.
Press the [ZERO/TEST] key.
The "Range" symbol flashes for approx. 8 seconds.
The display shows the following in alternating mode:
Perform measurement with a standard of known concentration as
described.
Press the [ZERO/TEST] key.
The "Range" symbol flashes for approx. 3 seconds.
are
21

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Md100 codMd100 dcoMd100 doc

Table des Matières