Calibration Mode; User Calibration - Lovibond TurbiCheck Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Calibration Mode

GB
0.10
StAn
Read
01:00
0.10
20
StAn
Read
01:00
20
200
StAn
Read
01:00
200
800
StAn
Read
TurbiCheck_1b 01/2011

User calibration

4
After confirming the selection with the [MODE] key 0.10/StAn appears
alternating in the display.
Place the < 0.1 NTU standard in the sample chamber making sure that
the marks are aligned. Place the cover on the instrument.
Press the [READ] key.
The countdown starts.
After the countdown is finished the reading starts automatically.
The display flashes for approx. 8 seconds.
After the measurement 20/StAn appears alternating in the display.
Invert the 20 NTU standard and place it in the sample chamber,
making sure that the marks are aligned. Place the cover on the
instrument.
Press the [READ] key.
The countdown starts.
After the countdown is finished the reading starts automatically.
The display flashes for approx. 8 seconds.
After the measurement 200/StAn appears alternating in the display.
Invert the 200 NTU standard and place it in the sample chamber,
making sure that the marks are aligned. Place the cover on the
instrument.
Press the [READ] key.
The countdown starts.
After the countdown is finished the reading starts automatically.
The display flashes for approx. 8 seconds.
After the measurement 800/StAn appears alternating in the display.
Invert the 800 NTU standard and place it in the sample chamber,
making sure that the marks are aligned. Place the cover on the
instrument.
Press the [READ] key.
33

Publicité

Table des Matières
loading

Table des Matières