Télécharger Imprimer la page

Agilent Technologies Turbo-V 2300 Rack Controller Notice De Mode D'emploi page 215

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
Figure 11
All the logic output from the controller must be connected at P2
mating connector. With the optional P2 mating connector make the
connection with AWG 24 (0.25 mm
indicated in the figure to obtain the desired capability. The following
table describes the signals available on the connector.
The following figure shows a typical logic output connection (relay
coil) but any other device may be connected e.g. a LED, a computer,
etc., and the related simplified circuit of the controller.
Figure 12
Tab. 4
PIN
2-1
4-11
5-12
6-13
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)
Logic output connector
Typical output connection open collector
DESCRIPTION
Analog output Voltage of power drawn by the turbopump (pin 2 positive, pin 1
negative). 1 Vdc proportional to 100 W.
R1 signal 24 V, 60 mA optically isolated output (pin 11 positive, pin 4 negative).
The output Voltage will be present when the rotational speed of the pump is
higher than the selected speed threshold
LOW SPEED signal, 24 V, 60 mA, optically isolated output (pin 12 positive, pin 5
negative). The output Voltage will be present when the low speed mode is
selected, either through, the front panel, the remote segnal, or RS 232/RS 485.
START signal 24 V, 60 mA, optically isolated output (pin 13 positive, pin 6
negative). The output Voltage will be present when the START push-button on
Technical Information
2
) or smaller wire to the pins
8
8
15
15
9
9
1
1
17
Interconnections
+ 24
215/262

Publicité

loading

Ce manuel est également adapté pour:

96995399699540