Heater Band Installation - Agilent Technologies TV 301 Mode D'emploi

Table des Matières

Publicité

17
Technical Information

Heater Band Installation

Heater Band Installation

Figure 33
The heater band model 9699804 and 9699803 can be used to heat the
pump casing when a bakeout is needed.
The heater band is applied to the upper part of the pump casing, as
shown in the figure, and heats it to a temperature of about 80 °C.
The heater band must be mounted such that there is perfect thermal
contact with the pump wall to obtain fast and efficient heating.
Figure 34
256/284
TV 301 Navigator User Manual / 87-900-946-01 (N)