Agilent Technologies TV 301 Mode D'emploi page 213

Table des Matières

Publicité

The TV 301 Navigator can be installed in any position. Fix the TV 301
Navigator in a stable position connecting the inlet flange of the
turbopump to a fixed counter-flange capable of withstanding a torque
of 1000 Nm around its axis.
For example the ISO 100 K flange can be fixed using high strength
steel clamps (as Agilent model IC 63250 DCMZ).
The following table shows, for each flange dimension, the necessary
number of clamps IC 63250 DCMZ and the relevant fixing torque.
Tab. 2
FLANGE
ISO 100 K
ISO 160 K
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
The TV 301 Navigator cannot be fixed by means of its base.
NOTE
The TV 301 Navigator belongs to the second installation (or overvoltage)
CAUTION!
category as per directive EN 61010-1. Connect the device to a mains line that
satisfy the above category. The TV 301 Navigator has Input/Output and serial
communication connectors that must be connected to external circuits in such a
way that no electrical part is accessible. Be sure that the insulation of the device
connected to the TV 301 Navigator is adequate even in the case of single fault as
per directive EN 61010-1.
For installation of optional accessories, see "Technical Information".
TV 301 Navigator User Manual / 87-900-946-01
FIXING DEVICE
M10 clamps
M10 clamps
(N)
Instructions for Use
Installation
N.
FIXING TORQUE
4
22 Nm
4
22 Nm
213/284
16

Publicité

Table des Matières
loading

Table des Matières