Agilent Technologies TV 301 Mode D'emploi page 251

Table des Matières

Publicité

N.
203
204
205
206
210
300
301
302
320
to
399
400
402
404
500
503
504
TV 301 Navigator User Manual / 87-900-946-01 (N)
Read/Write
Datatype
R
N
R
N
R
N
R
N
R (see note
N
3.)
R
N
R
N
R
N
Reserved to Agilent service
R
A
R
A
R
A
Reserved to Agilent service
R/W
N
R/W
L
Technical Information
RS 232/RS 485 communication description
Description
Driving frequency in Hz
Pump temperature in °C
Pump status
Error code
Actual rotation speed in Hz
Cycle time in minutes (zeroed
by the reset command)
Cycle number (zeroed by the
reset command)
Pump life in hours (zeroed by
the reset command)
CRC EPROM (QE)
CRC Param. (PA)
CRC Parameter structure
RS 485 address
Serial type select
17
Admitted Values
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description:
see the
following figure
100 to 963
0 to 999999
0 to 9999
0 to 999999
QE8XXXX
(where "XXXX"
are variable)
PA8XXXX
(where "XXXX"
are variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)
251/284