Hazardous Locations
Specifications
Model
Temperature
Electrical Connections
Ingress Protection
Electromagnetic Compatibility
Dielectric Strength Conformance LCD
Display
Manufacturer
Approved Battery
Replace battery only with Ralston Instruments P/N: DGAU-0185 Lithium-Ion Battery Pack
Nominal Voltage: 3.75 Volts
Rated Capacity: 4.8 Ah
Standards
This product complies with the following standards. Refer to EC declaration of Conformity for
specific details.
ATEX
IECEx
30
FLP1 HAZARDOUS LOCATION INSTRUCTIONS
Normas e aprovações de
agências ..........................................43
Norme e approvazioni
delle agenzie ...................................45
Standards und behördliche
Genehmigungen ..............................48
FLP1
Operating: -10 to 50°C, Process Media: -10 to 50°C,
Storage: -40 to 75 °C
USB Connection Um=5.25V, External Power Um=21V
IP67 (With both USB port and charging port cover
installed)
In the presence of an RF field ≥ 3 V/m on all pressure
ranges an acceptable level of reduction in performance
may be experienced
Conforms to the dielectric strength requirements of
IEC 60079-0
Ralston Instruments
15035 Cross Creek Parkway
Newbury, OH 44065 USA
www.ralstoninst.com
Charging Temperature Range: -20°C to 60°C
Discharging Temperature Range: -10°C to 50°C
Ex II 1 G Ex ia IIC T4 Ga
-10 °C ≤ Ta ≤ 50 °C IP67
Ex ia IIC T4 Ga
-10 °C ≤ Ta ≤ 50 °C IP67
P/N DGAU-0323 REV 0 8/3/2015