TROUBLESHOOTING GUIDE
Problem
"No Disk, Test Aborted"
"No Disk Cover, Test
Aborted"
"Lid Open, Test Aborted"
"Used Disk"
Meter Check Disk
"Source" appears on
test screen instead of
"Sanitizer"
!
on Test History
screen
Range Error
Output Error
Tap to select the market type. Select Pool/Spa. Tap
menu.
About...lists the Serial Number, Firmware Version, Bluetooth MAC address,
Bluetooth Version and Test Count. The Test Count shows the number of
complete tests that have been performed over the lifetime of the meter. Tap
to return to the Settings menu.
Ranges Enabled allows the option of having test results that are out of range
of the reagent system displayed in red. The default setting is off .
When Show Market Type is selected, the market type will be displayed on the
Testing Screen. The default setting is OFF. Tap
menu.
Select Disk Detection to display an error message and abort the test when
improper testing conditions, such as No Disk, No Disk Cover, Used Disk or
Meter Check Disk, are present that would result in inaccurate or no test
results. The default setting is OFF. Tap
Reason
No disk in chamber
Empty chamber
Disk cover was not used
Lid opened
Reacted disk in chamber
Meter Check Disk in chamber
instead of reagent disk
Wrong Market Type was chosen
Meter is using the default blank
due to under fi lled disk or air
bubble.
Raw data out of range
Decreased light intensity.
Possibly dirty lens
to return to the Settings menu.
Solution
Place fi lled disk in chamber. Cover
with disk cover. Select OK. Start test.
Insert fi lled disk. Cover with disk
cover. Select OK. Start test.
Use disk cover. Select OK. Start test.
Close lid. Select OK. Start test.
Select "Continue" to go to Test
Results screen. Select "Abort" to go
to Testing screen and run test with
new disk.
Select "Continue" to go to Test
Results screen. Select "Abort" to go
to Testing screen and run test with
reagent disk.
Go to Settings. Change Market Type
to Pool/Spa
Solution: Fill disk correctly (see
FILLING)
Contact Support
Clean lens (see CLEANING). Follow
Range Check Procedure. If error
message persists, contact Support.