VI.5 - Sterilisation
• The instruments are to be sterilised before first use and after each use.
• Only sterilise cleaned, lubricated and tested instruments.
• We recommend sterilisation that complies with the EN 13060 standard,
class B. All other methods are forbidden.
• Independently put each MD in a sterilisation sachet that complies with
current standards and the size of the MD.
• Comply with the space between the sachets and do not overload the
autoclave.
• These MDs and their accessories (apart from accessories with the
g
logo
W
which are single use) must be sterilised at 135°C for 18 minutes
minimum (sterilisation time).
x
• Check that the device is completely dry at the end of the cycle.
VI.6 - Storage
D
Keep the MD in a sterilisation sachet in a clean and dry place.
VII - REPAIR
-----------------------------------------------------------------------------------------------
In the event of breakdown, please contact your approved distributor or
your local B.A. International repair centre.
Repairs must only be carried out by an approved repairer or by B.A.
E
International After Sales Department, only with B.A. International
original replacement parts.
For all revisions or repairs, the MD must be returned complete and sterile
with proof of sterility. It must be accompanied by a document outlining the
problem and showing the complete contact details of the user.
The replacement of removable parts is covered for 7 years.