Electromagnetic Compatibility - ADE M320600-01 Guide D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 41
Medical electrical equipment is subject to special precautions regarding EMC and must be installed
and commissioned in accordance with the guidance below.
Portable and mobile HF devices (e.g. mobile phones) may affect medical electrical equipment.
The use of third-party accessories may increase the emission or reduce the immunity of the device.
Do not use mobile phones or similar devices that emit electromagnetic fields near the product. This
could adversely affect the functionality of the product.
Guidance and manufacturer's declaration - electromagnetic emissions
This PRODUCT is intended for use in the electromagnetic environment specified below. The customer or the user of
the PRODUCT should make sure that it is used in such an environment.
Emission tests
HF emissions as per
CISPR 11/EN55011
HF emissions as per
CISPR 11/EN55011
Harmonic emissions
as per IEC 61000-3-2
Voltage fluctuations/
flicker emissions as
per IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
This PRODUCT is intended for use in the electromagnetic environment specified below. The customer or the user of
the PRODUCT should make sure that it is used in such an environment.
Immunity tests
Electrostatic discharge
(ESD) according to
IEC 61000-4-2
Electrical fast
transient/
burst as per
IEC 61000-4-4
Surges according to
IEC 61000-4-5
Voltage drops, short-
term interruptions, and
fluctuations of the
supply voltage
according to
IEC 61000-4-11
M320600-01+02-210429-Rev004-UM-en
11.

Electromagnetic compatibility

Compliance
Group 2
Class B
Class A
Complies
IEC 60601 Test Level
± 8 kV contact discharge
± 15 kV air discharge
± 2 kV power lines
± 1 kV for input and
output lines
± 1 kV conductor-
conductor
± 2 kV conductor-earth
0% UT for 1/2 period
(100% break-in)
0% UT for 1 period
(100% break-in)
40% UT for 5 periods
(60% break-in)
70% UT for 25 periods
(30% break-in)
80% UT for 250 periods
(20% break-in)
0% UT for 250 periods
(short interruption)
Electromagnetic environment - guidance
The device must emit electromagnetic energy to perform its
intended function. Electronic devices in the area can be affected.
The PRODUCT is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage network that supplies buildings used for
domestic purposes.
Compliance level
± 8 kV contact discharge
± 15 kV air discharge
± 2 kV power lines
± 1 kV for input and
output lines
± 1 kV conductor-
conductor
± 2 kV conductor-earth
0% UT for 1/2 period
(100% break-in)
0% UT for 1 period
(100% break-in)
40% UT for 5 periods
(60% break-in)
70% UT for 25 periods
(30% break-in)
80% UT for 250 periods
(20% break-in)
0% UT for 250 periods
(short interruption)
Electromagnetic environment -
guidance
Floors should be made of wood or
concrete, or covered with ceramic tiles.
If the floor is covered with synthetic
material, the relative air humidity must
be at least 30%.
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
If the user of the PRODUCT requires
continued operation even during power
interruptions, it is recommended that
the PRODUCT be supplied by an
uninterruptible power source or a
battery.
Page 37 of 120

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

M320600-02

Table des Matières