Troubleshooting Guide - Gram AM Série Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TROUbLEShOOTINg gUIDE

Sympton
The balance does not turn on
The balance turn on but
the zero indicators doesn't
appear
The balance doesn't weigh in
whole capacity
The balance doesn't operate
correctly
The balance goes Off by
itself
The display is full of zeros
14
Cause / Solution
Discharged battery / Charge it using the AC/
Adaptor
Plug the AC/Adaptor.
Connect the power connector to the balance.
The weighing pan is not installed correctly.
There is a possible object on the weighing pan.
Remove all items from the pan and connect again
the balance.
Tare indicator activated. Disconnect the balance,
remove all items from the pan and connect the
balance again.
An object is located between the weighing pan and
the balance. Remove this object.
There are air currents, turbulences or
electromagnetic interferences. Place the balance
away from these disturbances.
You are using an incorrect weight unit. Check in
display that weight unit is the correct, otherwise,
press the MODE key to select it.
The balance is not placed on a flat surface or is
not levelled. Adjust the adjustable feet until the
level bubble remains in the centre of the circle.
There are air currents, turbulences or
electromagnetic interferences. Place the balance
away from this disturbances.
The balance needs to be calibrated. Calibrate it or
contact with your Distributor.
Auto-power off function activated. Deactivate this
function.
The battery indicator is on. Recharge the battery
with the AC / Adaptor.
The object located on the weighing pan exceeds
the maximum capacity of the balance. Remove the
object from weighing pan and try again.