Varian Turbo-V 300HT Notice De Mode D'emploi page 68

Masquer les pouces Voir aussi pour Turbo-V 300HT:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
WIN
R W T
000
X
X
L
START/STOP
001
X
X
L
LOW SPEED [0=OFF / 1=ON]
100
X
X
L SOFT START [0=NO / 1=YES]
101
X
X
L DEAT TIME [0=NO / 1=YES]
102
X
X
L WATER COOLING [0=NO / 1=YES]
103
X
X
N SPEED THRESHOLD
104
X
X
N RUN UP TIME (0÷359.999 sec)
106
X
X
N SPEED ADJUST
107
X
X
N MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
108
X
X
N BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
109
X
L PUMP LIFE RESET (TYPE "ON" TO RESET)
200
X
N CURRENT [ A]
201
X
N VOLTAGE [ V]
202
X
N POWER [ W]
203
X
N FREQUENCY [ Krpm]
204
X
N TEMPERATURE (0÷99) [ °C]
205
X
N PUMP STATE (0÷6) [STOP, WAITING INTERLOCK, STARTING, NORMAL, HIGH LOAD,
FAILURE, APPROACHING]
206
X
N ERROR CODE (0÷7) [NO ERROR, OVERVOLTAGE, SHORT CIRCUIT,CHECK CONN.,
TOO HIGH LOAD, OVERRIDE, OVERTEMP PUMP, OVERTEMP CONTR.]
207
X
L STATE R1 [0=OFF / 1=ON]
208
X
L STATE R2 [0=OFF / 1=ON]
300
X
N
CYCLE TIME
301
X
N CYCLE NUMBER
302
X
N PUMP LIFE
400
X
A
CRC PROGRAM LISTING
402
X
A
CRC PARAMETER LISTING
WIN = Window
R
= Read
W
= Write
(0-999.999) IN MINUTES
(0- 65.535) IN COUNTS
(0-999.999) IN HOURS
T = Type:
L = Logical
N = Numeric
A = Alphanumeric
62
TECHNICAL INFORMATION
Description
87-900-884-01 (C)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9870

Table des Matières