Maintenance And Cleaning - IKA VC 10 Mode D'emploi

Masquer les pouces Voir aussi pour VC 10:
Table des Matières

Publicité

PC 1.1 Cable (Device to PC)
Required for connecting the 9-pin socket to a PC.
1
6
2
7
3
8
4
9
5
1
RxD 2
TxD 3
4
GND 5
6
RTS 7
CTS 8
9
Connection VC 10 - PC
or
Mikro B
Connection cable MVP 10.100
This cable is required to connect the VC 10 to the IKA
The instrument is maintenance-free. It is only subject to the natu-
ral wear and tear of components and their statistical failure rate.
For cleaning disconnect the main plug!
Use only cleaning agents which have been approved by IKA
clean IKA
®
instruments.
Dirt
Dyes
Construction materials
Cosmetics
Foodstuffs
Fuel
Not listed materials
Wear protective gloves during cleaning the instruments.
Electrical instruments may not be placed in the cleansing
agent for the purpose of cleaning.
USB A
®

Maintenance and cleaning

Cleaning agent
isopropyl alcohol
water containing tenside, isopro-
pyl alcohol
water containing tenside,
isopropyl alcohol
water containing tenside
water containing tenside
please request information from
IKA
®
application suppor
9-pin RS 232
USB Mikro B
MVP 10 digital membrane vacuum pump (speed mode).
Do not allow moisture to get into the instrument when
cleaning.
Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with
IKA
that this method does not destroy the instrument.
®
to
®
Spare parts order
When ordering spare parts, please give:
machine type,
manufacturing number, see type plate,
item and designation of the spare part see www.ika.com,
spare parts diagram and spare parts list.
software version
Repair
Please send in instrument for repair only after it has been
cleaned and is free from any materials which may consti-
tute a health hazard.
For this, you should request the "Decontamination Certifica-
te" from IKA
®
website www.ika.com.
If you require servicing, return the instrument in its original pa-
ckaging. Storage packaging is not sufficient. Please also use suita-
ble transport packaging.
9
8
7
6
1
2 RxD
3 TxD
4
5 GND
6
7 RTS
8 CTS
9
, or use the download printout of it from the IKA
5
4
3
2
1
PC
Fig. 10
9-pin RS 232
USB A
Fig. 11
Fig. 12
®
47

Publicité

Table des Matières
loading

Table des Matières