Knick Stratos Pro A2 Oxy Série Guide Rapide page 16

Masquer les pouces Voir aussi pour Stratos Pro A2 Oxy Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 33
Error Messages
Info text
(is displayed in case of
Error
fault when the Info key is
pressed)
* Memosens or ISM sensors
16
Problem
Possible causes
Failure in sensor *
Error in cal data *
Display range violation
SAT saturation
CONC concentraton or
GAS volume concentration
Measuring range of
sensor exceeded
Temperature range
violation
Sensocheck
Load error
Output current 1
< 3.8 mA
Output current 1
> 20.5 mA
Output current 2
< 3.8 mA
Output current 2
> 20.5 mA