DIRECTIONS FOR USE

Press the ON / OFF button (4). The display (3) is activated to show all
1.
segments for 2 seconds.
2.
The last measurement reading will automatically be shown on the display
for 2 seconds with the «M» icon (7).
3.
When the «°C» icon is flashing, a beep sound is heard and the thermometer
is ready for measuring (8).
FOREHEAD MEASUREMENT
4.
5.
Place the probe gently onto the area approximately 1 cm above the center
of the eyebrow. If the eyebrow area is covered with hair, sweat or dirt,
please remove the obstacle beforehand to improve the reading accuracy.
Press the START button (2) once and release it. Start to move the device
6.
steadily from 1 cm above the center of the eyebrow to the temple area.
A long beep sound will emit after 3 seconds to acknowledge the comple-
tion of measurement. If the probe movement is too slow to reach the temple
area before the long beep sound is heard, repeat the measurement as
described above, but simply move the probe a little quicker this time.
7.
Read the recorded temperature from the LCD display.
NOTE:
Patients and thermometer should stay in similar room condition for at
least 30 minutes.
Do not take a measurement while or immediately after nursing a baby.
Do not use the thermometer in high humidity environments.
Patients should not drink, eat or exercise before / while taking the measure-
ment.
Before placing the thermometer probe onto the measurement area,
remove dirt, hair, or sweat.
Do not move the measurement device from the measuring area before
hearing the termination beep.
Always take the temperature in the same location, since temperature
readings may vary according to locations.
10 short beeps and a red LCD backlight alert the patient that he / she may
have a temperature equal to or higher than 37.5 °C.