Endress+Hauser Liquisys M COM223 Manuel De Mise En Service page 113

Transmetteur d'oxygène dissous
Masquer les pouces Voir aussi pour Liquisys M COM223:
Table des Matières

Publicité

Liquisys M COM223/253
Measured value
in hPa
only display switch!!
Selection of
measuring range
20 mg/l
60 mg/l
200%Sat 600% Sat
400 hPa
1200 hPa
A6
Feedforword control
to PID controller
Off; lin = linear
Basic = basic load
Z6
Selection of
value pair in table
1
1 ... no. of value pairs
assign
O333
Enter 20mA value
10.00 mg/l;
0...20 mg/l
100.0%SAT; 0...200%SAT
200 hPa;
0...400hPa
40.0 °C;
-10...60 °C
O313
Activate
alarm contact
yes; no
F6
Setting of maximum
permissible limit
violation period
(below limit)
480 min; 0...2000 min
P6
Endress+Hauser
Measured value
Measured value
Current input in %
Sensor signal in nA / mV
Feedforward control
gain = 1 at
50%
0 ... 100%
Z7
Entry of x-value
Entry of y-value
(measured value)
(current)
0.00 mg/l; 0...20 mg/l
0.0
; 0...200
%SAT
%SAT
0 hPa;
0...400 hPa
0.00mA; 0...20.00mA
O334
O335
Activate error current
Automatic start of
for previously set
cleaning function
error
no; yes
no; yes
F7
Setting of maximum
Setting of
permissible limit
limit value
violation period
5.00 mg/l;
0...20 mg/l
50.0 %SAT; 0...200%SAT
(upper limit)
200 hPa;
0...400 hPa
240 min; 0...2000 min
P7
Measured value
Current input in mA
Table status o.k.
yes; no
O336
Selection "next error"
or return to menu
Next=next errror
<--- R
F8
F9
P8
Annexe
Field for customer
settings
A0027487-FR
113