Annexe

CALIBRATION
C
Edit:
Read:
Code 22
any code
MEAS. VAL. DISPLAY
with
TEMPERATURE
DISPLAY
in °C
Edit:
Read:
Code 22
any code
SETUP 1
A
SETUP 2
B
CURRENT INPUT
Z
CURRENT OUTPUT
O
ALARM
F
CHECK
P
112
13
Annexe
Entry of
calibration value
Ref = entry of
reference value
current value
C1 (3)
Start
calibration
Countdown
600 s --> 0 s
H O = water
2
C1 (2)
Start
calibration
Countdown
600 s --> 0 s
on air
C1 (1)
Temperature display
in °F
1st error is displayed
(if present)
Selection of
Selection of
operating mode
display unit
mg/l % SAT hPa
mg/l ppm ppb
A1
Entry of
Entry of
salt concentration
process temperature
current meas. value
0.0 % 0.0 ... 4.0 %
-10.0 ... 60.0 °C
B1
Controller switch-off
Delay for controller
by current input
switch-off by current
input
Off; Input
0 s
0 ... 2000 s
Z1
Selection of
Select meas. unit for
current output
2nd current output
°C; mg/l; % Sat,
Out 1, Out 2
hPa; Contr
O1
Selection of
Selection of
contact type
alarm delay unit
Latch=latching
min; s
Momen=momentary
F1
Selection of
Alarm delay
alarm threshold
monitoring
0 min (s)
Off; Low; High;
0 ... 2000 min (s)
(depends on F2)
Lo
+Hi=low and high
P1
Slope is
Calibration status
displayed
is displayed
100.0 %
75.0 ... 140.0 %
o.k.
C131
C132
Store calibration
Calibration status
results
is displayed
yes; no; new
o.k.
Exxx
C121
C122
Calibration status
Store calibration
results
is displayed
o.k.
Exxx
yes; no; new
C111
C112
Measured value
Temperature display
in mg/l
suppressed
Other errors are
displayed
(up to max. 10 errors)
*
Entry of
altitude
------------
0 m
A2
A3
Display of
temp. difference
(Offset)
current offset
-5.0 ... +5.0 °C
B2
B3
Delay for controller
Switch-off limit value
switch-on by current
for current input
input
0 s
0 ... 2000 s
50%
Z2
Z3
Selection of
table option
read
Tab = Table
O3 (3)
Entry of
simulation value
current value
0 ... 22.00 mA
sim = Simulation
O3 (2)
Selection of
current range
4-20 mA; 0-20 mA
lin = linear
O2
O3 (1)
Alarm delay
Setting of
error current
0 min (s)
0 ... 2000 min (s)
22 mA; 2.4 mA
(depends on F2)
F2
F3
Setting of lower
Setting of upper
alarm threshold
alarm threshold
0.00 mg/l;
0...19 mg/l
20.00 mg/l; 1...20 mg/l
0.0%SAT;
0...190%SAT
200.0%
SAT
0 hPa;
0... 380 hPa
400 hPa;
P2
P3
Liquisys M COM223/253
Store calibration
results
Exxx
yes; no; new
C133
C134
Operating matrix
DX/DS with COS41
C123
C113
Measured value
in % Sat
Entry of measured
value damping
1 (no damping)
0 ... 4000 m
1 ... 60
A4
A5
*
these fields are not available with DX versions
Switch-off direction
for current input
0 ... 100%
Low; High
Z4
Z5
Enter number of
value pairs in table
edit
1
1 ... 10
O331
O332
O321
Enter 0/4mA value
0.00 mg/l; 0...20 mg/l
0.0%Sat; 0...200% Sat
0 hPa;
0...400 hPa
0.0 °C;
-10...60 °C
O312
O311
Selection of
error numberl
1
1 ... 255
F4
F5
Selection of process monitoring
off; AC; CC
monitoring
limit (AC):
;10...200%
SAT
AC+CC
±0.1 mg/l
20...400hPa
±2 hPa (in 1 h)
P4
±1 % SAT
P5
A0027486-FR