If you want to also perform the cleaning for DISINFECTION, jump to the DISINFECTION
paragraph.
After having sanitised the accessories, shake them vigorously and place them on a paper
towel. Alternatively, dry them with a jet of hot air (for example, a hair dryer).
DISINFECTION
After sanitising the nebuliser cup and the accessories, disinfect them choosing one of the
methods described below.
method A: Accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium
hypochlorite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: Disinfect the accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 by boiling them in
water for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Disinfect the accessories C1.1-C1.2-C1.3-C2-C3-C4-C5 with a hot steam
steriliser for baby-bottle (not the microwave type). Perform the process faithfully following
the instructions of the steriliser. To ensure that the disinfection is effective, choose a
steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away
from dust.

REPLACING THE FILTER

The device has a suction filter (A3) that should be replaced
when it is dirty or changes color. Do not wash or reuse the
same filter. Regular replacement of the filter is required to
help and ensure correct performance of the compressor.
Check the filter on a regular basis. Contact your local dealer
or authorized service center for spare filters.
To replace the filter take it out by pulling it as shown in figure.
The filter has been designed so that it is always fixed in its seat. Do not replace the filter
during use. Only use original accessories and spare parts by Flaem, we disclaim any
liability in the event of using non original spare parts or accessories.