Varian Turbo-V 70 Notice De Mode D'emploi page 61

Masquer les pouces Voir aussi pour Turbo-V 70:
Table des Matières

Publicité

The Soft Start mode may be deselected/selected only
when the pump is stopped.
If necessary press CYCLE NUMBER to select YES or
NO, then enter the selection by pressing the PUMP
CURRENT push-button. The display shows:
F R O N T
/
R E M O T E
S E L E C T
I O N : X X X X X X
where: XXXXXX = means the word FRONT or
REMOTE, or RS 232 depending on the last
selection.
After choosing the desired selection by pressing the
CYCLE push-button, press the PUMP CURRENT
push-button to enter the value.
The display shows:
R S 2 3 2
B A U D
R A T E
where: XXXX = means 600, 1200, 2400, 4800, 9600
baud
rate
for
the
communication.
NOTE
/
2 3 2
X X X X
host
computer
or
printer
If necessary, select the desired value by pressing the
CYCLE NUMBER, then enter the value by pressing the
PUMP CURRENT push-button.
The display shows:
R S 2 3 2
S E L E C T
where: XXXX == means HOST or PRINT.
Select HOST or PRINT by pressing the CYCLE
push-button.
With
bidirectional communication is established by selecting
HOST. Data are sent to an external computer every
time the external computer asks for the values. The
data available are:
Pump/controller operating condition
Cycle time
Pump life
Pump temperature
Pump current
Pump voltage
Controller output frequency
Cycle number
R1 condition
R2 condition
Life time and cycle # zeroing
Configuration parameter readings
Configuration parameter setting
If PRINT is selected and a printer is connected on RS
232 line, a unidirectional communication is established
and every minute the data are sent to the printer, even
if the pump is not running.
The set of data available are:
Pump speed KRPM
Pump temperature
Pump current A
Pump power W
R1 condition
R2 condition
55
TECHNICAL INFORMATION
H O S T
/
P R I
I O N :
X X X X X
the
RS
232
connected,
87-900-870-01 (B)
N T
a

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9506

Table des Matières