10. ERROR CODES

Any malfunction during operation may be identified by an error message on the display.
Proceed as follows in such cases:
① Switch off device using the power switch.
② Carry out corrective measures (see table below).
③ Restart the device.
Error code
Cause
Error 3
Inside temperature is too
high
Error 4
Motor cannot run normally
(cannot start or unable to
reach the target speed)
Error 47
Overload
Error 48
Error 50
Rotor is protected to
prevent overload
If the actions described fail to resolve the fault or another error code is displayed, then take the
following steps:
Contact the service department: support@vbdevices.com
If authorized by the service department, send the device for repair, including a short
description of the fault, according to the instructions.

11. MAINTENANCE

Maintenance
VARIXIO® Pro Mag is maintenance-free but will undergo normal wear.
Device exterior labelling 1 - Information.
Label present on the back of the equipment
next to the power connector, in accordance
with the requirements established in the CE
Directive
Effect
motor off
motor off
motor off
motor off
For cleaning disconnect the main plug.
Solution
- Switch off the device and let it cool
down.
- Restart the device.
- Switch off the device.
- Check the structure for possible
blockages.
- Switch off the device.
- Decrease the speed setting or the
load.
- Switch off the device.
EN