13. Maintenance and care

Check the emergency shower once a month to make sure that it is in working order.
(Guidelines for Laboratories BGI/GUV-I 850-0)
As a measure for the prevention of microbiological contamination, it is recom-
mended that you replace the fill water in the fitting at shorter intervals by opening the
valve.
Use suitable cleaning agents which do not attack the fitting in a proper manner and
rinse with water after use. High-pressure cleaners must not be used for cleaning.

14. Troubleshooting

Fault
Water does not flow
Manual actuator is
catching
Body shower is
dripping
If you cannot rectify a fault or if it is not in the list, please contact our customer
service department!

15. Spare parts

Description
Ball valve . . . . . . . . . . . . . . . . . . . 2030025022
Other spare parts on request; see the wall bracket for material numbers.
Cause
– Water supply interrupted
– Ball valve not correctly aligned
– Pull rod or manual actuator bent
– Ball valve stiff
– Ball valve defective
Order No.
Remedy
➯ Restore it
➯ Align it
➯ Straighten it
➯ Replace it
➯ Replace it