7.2 diagnostic display Indicators and Controls

Rinse Indicator
L ights when the Power
Level Control is turned
fully counterclockwise.
Rinse mode provides
lavage to flush the
procedural area with
negligible tip movement.
Service Indicator
Lights when the system is
not functioning properly.
This diplay has three
distinct modes.
• A fast blink (3 blinks per
second) indicates an
improper set-up.
• A slow blink (1 blink per
second) means the
system is operating out
of factory specifications.
• A steady light indicates
the system is overheating.
R efer to Section 10.1 for
Troubleshooting guidelines.
Blue Zone Indicator
L ights when the Power Level Control is positioned
in t he Blue Zone of the power scale.
Ideal for effective subgingival debridement and
g reater patient comfort.
Purge Control
L ights when the Purge function is activated.
T o activate Purge, remove insert from the handpiece,
and press the Purge button. Water will purge through
system for 2 minutes. For optimal efficiency, turn the
handpiece lavage control to maximum water flow. To
deactivate mode during the 2 minute cycle, press
Purge button again or press foot control.
The Purge Control is also used during the Foot Control
Synchronization process. See Section 6.8.
9
Boost Indicator
L ights when the
Boost Mode has
been activated.
Low Battery
Indicator
L ights when the foot
control battery power
is a pproaching end of
life. Replace batteries
as instructed in
Section 6.7.
Power Indicator
Lights (3 Sec. delay)
when the Main Power
ON/OFF Switch is ON
("I" position).