4. After cleaning/disinfecting, the transducer should be immersed
briefly in distilled water and left to dry.
5. Re-assemble the mouthpiece holder.
Housing
Turbine transducer
Cut-Out
Locating pip
CAUTION: Do not attempt to wash or immerse the MicroLab
transducer housing in water or cleaning fluid, as there are
electronic components inside that will be permanently damaged.

Servicing

There is no routine maintenance required for the MicroLab and
there are no user serviceable parts in this spirometer. Please
return the unit to Vyaire or an authorized agent, if servicing is
required.
If your spirometer requires service or repair please see page 101
for contact details.