Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroLab is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroLab should assure that it is used in
such an environment.
Immunity
Test
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
IMMUNITY
to proximity
fields from
RF wireless
communications
equipment IEC
61000-4-3
IEC 60601 test level
3 V
0,15 MHz – 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
28 V/m
450 MHz, 50% PM at 18 Hz
810 MHz, 50% PM at 18 Hz
870 MHz, 50% PM at 18 Hz
930 MHz, 50% PM at 18 Hz
1720 MHz, 50% PM at 217 Hz
1845 MHz, 50% PM at 217 Hz
1970 MHz, 50% PM at 217 Hz
2450 MHz, 50% PM at 217 Hz
27 V/m
385 MHz, 50% PM at 18 Hz
9 V/m
710 MHz, 50% PM at 217 Hz
745 MHz, 50% PM at 217 Hz
780 MHz, 50% PM at 217 Hz
5240 MHz, 50% PM at 217 Hz
5500 MHz, 50% PM at 217 Hz
5785 MHz, 50% PM at 217 Hz
46
Compliance Level
3 V
0,15 MHz – 80 MHz
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
28 V/m
450 MHz, 50% PM at 18 Hz
810 MHz, 50% PM at 18 Hz
870 MHz, 50% PM at 18 Hz
930 MHz, 50% PM at 18 Hz
1720 MHz, 50% PM at 217 Hz
1845 MHz, 50% PM at 217 Hz
1970 MHz, 50% PM at 217 Hz
2450 MHz, 50% PM at 217 Hz
27 V/m
385 MHz, 50% PM at 18 Hz
9 V/m
710 MHz, 50% PM at 217 Hz
745 MHz, 50% PM at 217 Hz
780 MHz, 50% PM at 217 Hz
5240 MHz, 50% PM at 217 Hz
5500 MHz, 50% PM at 217 Hz
5785 MHz, 50% PM at 217 Hz

Publicité