Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroLab is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroLab should assure that it is used
in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
IEC 60601 test
level
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
± 2 kV
100 kHz
repetition
frequency for
power supply
lines
± 0,5 kV, ± 1 kV,
± 2 kV
0 % U
; 0,5
T
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles
Single phase:
at 0°
and
0 % U
; 250/300
T
cycle
44
Compliance
Level
Contact: ± 8 kV
Air:
±2 kV, ± 4 kV, ±
8 kV, ± 15 kV
± 2 kV
100 kHz
repetition
frequency for
power supply
lines
± 0,5 kV, ± 1 kV,
± 2 kV
0 % U
; 0,5 cycle
T
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0 % U
; 1 cycle
T
and
70 % U
; 25/30
T
cycles
Single phase:
at 0°
and
0 % U
; 250/300
T
cycle
Electromagnetic
Environment
-Guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
Micro Lab requires
continued operation
during power main
interruptions, it is
recommended that
the Micro Lab be
powered from an
uninterruptible power
supply or a battery.