FORA IR20/IR21 is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below.
The customer or the user of FORA IR20/IR21 should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Manufacturer's declaration-electromagnetic emissions
Compliance
Group 1
FORA IR20/IR21 uses RF energy only for internal use. Therefore, its RF emissions are very
low and are not likely to cause any interference from nearby electronic equipment.
Class B
FORA IR20/IR21 is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Not applicable
Not applicable
Electromagnetic environment-guidance
(for home healthcare and professional healthcare)
EN-17