ASSEMBLY INSTRUCTIONS

This 2200KL scale has been calibrated at the factory. It is shipped in two assembled parts,
the weighing tray and the base assembly. Remove each assembly from its carton and unwrap
packing material carefully to prevent scratching the unit's parts.
1.
Place the base assembly on a steady surface.
2.
Attach weighing tray to the base assembly.
The scale may be operated by an AC power source adapter or its internal rechargeable
battery.
The scale will automatically switch to battery operation when an AC power source is absent.

SET UP

1.
Plug the scale's AC adapter into the back of the scale and into the power source.
2.
Press the ON/OFF button to turn the scale on. The display will show "start" and then
"0-00.0".
3.
When the self-test function is complete:
The digital display should read
The CHARGE indicator should be illuminated on the display.
"LB" should be indicated on the right side of the display. (To change the default
units of the scale to kilograms (Kg), please contact our Technical Support Center
at 1.800.638.3722 or 1.708.598.9100).
4.
Place a weight [not to exceed 15 Kg (30 Lb)] on the scale. The digital display should
------
blink "
" until the weight calculation is complete.
------
5.
Press the LBS/KG button to select the weight mode, either in kilograms or
pounds/ounces.
6.
Press the REWEIGH button; the digital display should read again the weight of the item
on the scale.
7.
Remove the weight from the scale. The digital display should continue to read the
weight of the item that was on the scale.
8.
Press the ZERO button; the digital display should return to zero.
9.
Disconnect the scale from the AC power source. The scale should automatically switch
to its internal battery and the CHARGE indicator should turn off.
NOTES:
1.
If the set up procedure failed, go to the troubleshooting instructions. If the problem is
not corrected, refer to qualified service personnel.
2.
At the end of set up procedure, connect the scale to an AC power source for at least
8 hours for battery recharging.
Figure 1. Assembly After Shipment
"0-00.0".