Cleaning the Easy Clean
Cleaning steps:
• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Soak the Easy Clean in the cleaning solution for, as a minimum, the time and the concentration
specified by the manufacturer of the cleaning agent (ENZOL, 2 %, 10 min).
• Brush the lumen and surface with a suitable lumen / surface brush until all visible soiling is removed.
• Rinse through the lumina with the cleaning solution (ENZOL, 2 %), at least 5 times using a disposable
syringe (min. volume 50 ml).
• Submerge the Easy Clean completely in deionized water.
• Rinse through the lumina with deionized water at least 3 times using a disposable syringe (min.
volume 50 ml).
• Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible on the
product.
• Dry the Easy Clean at room temperature.
• Place the Easy Clean in a sterilization bag.
The Easy Clean can be cleaned and sterilized in a fractional pre vacuum cycle at 132°C for 4
minutes
56
FB-543_HS_rev_A_01_Handy 3 Perio & Handy 3 PLUS.indd 56