channel. The bottom status line shows battery capacity, logger mode and alarm
status.
The
green
LED
flashes
The red LED is used to display limit alarms or status messages (battery change
...etc.). The logger also has an internal buzzer that supports the user interface.
This product is exclusively intended for the field of application described above.
It should only be used as described within these instructions.Unauthorized repairs,
modifications or changes to the product are prohibited and void any warranty!

5. How to use device

5.1 Device description

Fig. A (see page 2)
1. USB cover

2. USB-connector

3. LCD display see Fig. B
4. LED: green/red
5. Sensor compartment
Fig. B (see page 2)
1. Units of measurement line 1
2. Measurement line 1
3. Units of measurement line 2
4. Measurement line 2
5. Line3 for indication and
extrema measurements
EXT* = external probe (only visible
when external probe is connected) *TC*
AVG = average value,
Min = minimum value,
Max = maximum value
(no symbol) = current measurement value
If the display has been deactivated (display off via Software LogConnect), the battery symbol
and the symbol for recording (►) or configuration (II) are still active in Line 4 (status line).
20
every
30
seconds
during
recording.
(position of the sensor)
6. Battery case
7. Mode button,
8. Start / Stop button
9. lockable wall holder
10. External probes (TC only)
11. External probe (E only)
6. Line4 status line (from left to right):
Battery indication,
USB-connection indication,
Data logger is recording,
Data logger has been
configured by LogConnect
and is waiting for action,
Data logger has been
stopped and
OK, (without
► Symbol)
Data logger has been
stopped and X NOT OK
(e.g. Alarm level reached
(without ► Symbol)

5.2 Device start-up

ATTENTION! Please observe our battery recommendation strictly. Use only bat-
tery type LS 14250 3.6 volt of manufacturer SAFT or DYNAMIS Lithium Batt. LI-110
1/2 AA/S, respectively only batteries authorized by the manufacturer.
For operation take out the instrument from the packaging, remove the
display foil. The logger is already preset and ready for start. It can be
used immediately without any software!
By pressing any button or moving the instrument before
first operation the instrument displays FS (factory setting)
for 2 seconds, afterwards measurements are displayed for
2 minutes. Then the instrument display switch off. Repeated key hit or movement
reactivates the display.

5.2.1 Factory settings

Note the following default settings of the data logger before first use. By using the
LogConnect* software, the setting parameter can easily be changed:
Recording Interval:
5 min. LOG210/ LOG220, / LOG200 TC / LOG210 TC / LOG220 E
15 min. LOG200 / LOG200 E
Measuring interval:
During recording the measurement interval and recording
interval is the same! If the logger has been not started (NOT
RECORDING) the measuring interval is every 6 seconds for
15 minutes, afterwards the measuring interval is every 15 min.
for 24 hours, afterwards the measuring interval is once per
hour. If you press any button or move the device it will start
again to measure each 6 seconds.
Start possible by:
Key press
Stop possible by:
USB connect
Alarm:
off
Alarm delay:
0 s
Show measurements on display:
on
Power-Save mode for display:
on
21