Direct Healthcare Group SystemRoMedic Eva400 Série Manuel D'utilisation page 19

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 123
Lifting speed
Batteries
Internal charger
Motor (mast)
Motor (base)
Sound level
Material
Emergency lowering
Castors
Weight
IP class
Expected lifetime
Expected lifetime
of accessories
Operating forces
buttons on hand control
Turning diameter
Max. load
Operating environment
Storage and transportation
environment
System
I F U
Eva400
34 mm/s without load.
Two 12V, 2.9 Ah valve-regulated, sealed, lead accumulator (gel-type batteries)
Number of cycles when fully loaded: 300
External: Linak CH01
(9000368) Mains voltage
100-240VAC (50-60 Hz). Max.
400mA
DC 24V, 5 A. IP X4.
Operationtime: 10% at maxi-
mum continous running of 2
minutes, maximum 5 switching
cycles per minute. Push:
6000N.
24V, 5 A, IP X4. Operationtime:
10% at maximum continous
operation of 2 minutes,
maximum 5 cycles per minute.
Push: 3000N.
With load: upwards: 43 dB(A)
downwards: 44 dB(A).
Aluminum
Manual and electrical
Front 4", 100 mm, back 4", 100mm
74.9 lbs, 34 kg
IP X4
10 years
Consult the manual or information sheets regarding the respective accessories
4 N
1340 mm/ 54 inch
400 lbs, 180 kg
5 °C to 40°C, relative humidity 20% to 80% non-condensing, and atmospheric
pressure 700 to 1060 hPa.
-10 °C to + 50 °C, relative humidity 20% to 80% non-condensing, and atmospheric
Environment pressure 700 to 1060 hPa.
Internal: Included cable. Max. 400mA
DC 24 V, 10 A. IP X4. Operationtime: 10% at
maximum continous operation of 2 minutes, maxi-
mum 5.5 cycles per minute. Push: 10 000N.
24 V, 3 A, IP X4. Operationtime: 10% at maximum
continous operation of 2 minutes, maximum 5.5
cycles per minute. Push: 3000N.
With load: upwards: 74.7 dB(A) downwards: 52.6
dB(A).
450 lbs, 205 kg
Eva600
77.1 lbs, 35 kg
600 lbs, 270 kg
19

Publicité

Table des Matières
loading

Produits Connexes pour Direct Healthcare Group SystemRoMedic Eva400 Série

Table des Matières