Serial Communication Windows

WIN
TYPE
R W
000
L
X
X START/STOP (1= START ; 0= STOP)
001
L
X
X SPEED SELECTION ACTIVE [0=HS / 1=LS]
008
L
X
X REMOTE/SERIAL Configuration (1= Remote ; 0= Serial)
100
L
X
X SOFT START YES/NO (1= YES ; 0= NO)
101
N
X
X SET POINT TYPE (0 =Frequency; 1=Current; 2=Time)
102
N
X
X SET POINT VALUE [Hz, Ma, SEC]
103
N
X
X SET POINT MASK TIME [s]
104
L
X
X OUTPUT STATE (0= OFF if Val > threshold; 1 = ON if Val > threshold Default = 0
105
N
X
X HYSTERESIS IN THRESHOLD % [0-100]
107
L
X
X ACTIVE STOP (0=NO; 1=YES)
108
N
X
X BAUD RATE (3-4) [4800-9600]
109
L
X PUMP LIFE RESET [Write "1" to Reset]
111
N
X
X SET ANALOG OUTPUT TYPE [0=Power; 1=Current]
120
N
X
X SET ROTATIONAL FREQUENCY [Hz]
121
N
X
X MAX SETTABLE ROTATIONAL FREQUENCY [Hz] F<=F_MAX_ABS
127
N
X
200
N
X
201
N
X
202
N
X
203
N
X
204
N
X
205
N
X
206
N
X
211
N
X
216
N
X
300
N
X
301
N
X
302
N
X
319
A
X
320
A
X
321
A
X
323
A
X
325
A
X
400
A
X
401
A
X
402
A
X
404
A
X
406
A
X
407
A
X
500
L
X MONITOR MODE
WIN = Window
R
= Read
W
= Write
RUN UP TIME (8 – 60 minute)
CURRENT [mA]
VOLTAGE[V]
POWER [W]
DRIVING FREQUENCY [Hz]
PUMP TEMPERATURE [°C]
STATUS [0=stop; 1=interlock; 2=ramp; 3=regulation; 4=brake; 5=normal; 6=failure]
ERROR CODE:
Too high load
Short circuit
SoftStart Not Ended
Normal Not Reached
CONTROLLER SENSOR TEMPERATURE READING [208= 25°C - 128= 60°C]
AMBIENT SENSOR TEMPERATURE READING
CYCLE TIME [min]
CYCLE NUMBER
PUMP LIFE [h]
Controller Model
Base Pump Model Number (8 characters)
Modified Standard Model Number (4 characters)
Controller Serial Number (5 characters)
Electrical Modification Level (10 characters)
CRC PROGRAM LISTING [QE7xxxx]
CRC BOOTLOADER [BL1xxxx]
CRC PARAMETER LISTING [PA7xxxx]
CRC FILE PARAMETER STRUCTURE
PROGRAM LISTING CODE & REVISION
PARAMETER LISTING CODE & REVISION
L = Logical
N = Numeric
A = Alphanumeric
Description
1100 Hz < = F_imp < = FMAX
7
60
TECHNICAL INFORMATION
Default = 0
Default = 0
Default = 1117 Hz
Default = 0
Default = 2
Default = 0
Default = 4
Default = 0
0
No connection
Pump overtemp
Controller overtemp
Power fail
87-900-121-01 (B)