CardiAid is subject to maintenance periodically and after each use, as described below. Previously
performed maintenances, if there is any, can be tracked from the maintenance label on the device.
Periodical Maintenance Label
Danger!
Inspection, repair and other maintenance actions can be performed only by Cardia
International or a Cardia International authorized service provider. Do not attempt to
unscrew the device. This invalidates the warranty, may cause irreversible damage to the device and/
or serious injuries.

10.1 After-Use Maintenance

CardiAid should be subject to maintenance by Cardia International or a Cardia International authorized
service provider after every use. This ensures that CardiAid is in good condition and ready to use when
needed again. During this maintenance, main battery and electrodes are replaced, the incident data is
obtained from the device and some function tests are performed. Also, the emergency kit is replaced if
used. The next periodical maintenance should be performed at regular time. (See Section 10.2)

10.2 Periodical Maintenance

2nd and 4th Year Periodical Maintenance:
CardiAid should be subject to periodical maintenance at the end of 2nd and 4th years after purchase.
Date of next maintenance is indicated on the battery life label on backside of the device. During this
maintenance, main battery and electrodes are replaced; and some function tests are performed.
Caution!
The maintenances should be performed no later than the date on the battery life label. The
device cannot be used unless the maintenances are performed on time.
In some countries like Germany, legal period of Technical Safety Check (TSC) is 2 years according to
Medical Products Operation Regulation (Article 6). In these countries, CardiAid should be subjected to
TSC during 2nd and 4th year periodical maintenance.
6th Year Maintenance:
In order to be able to continue using the device after 6th year, CardiAid should be subjected to a
comprehensive maintenance at the end of 6th year after purchase. In addition to the periodical
maintenance, the status of technical components is checked. This maintenance is performed at the
factory. Please contact Cardia International for details.
After-use Maintenance Label

10. Maintenance