Inspection, repair and other maintenance actions can be performed only by Cardia
International or a Cardia International authorized service provider. Do not attempt to
unscrew the device. This invalidates the warranty, may cause serious injuries and/or irreversible
damage on the device.
Failure Messages of CardiAid
Visual
Battery Symbol and OK
Symbol are flashing
during stand-by.
Battery Symbol and
OK Symbol are lighting
continuously during
operation.
Battery Symbol and
Repair Symbol are
flashing during stand-by.
Battery Symbol and
Repair Symbol are
lighting continuously
during operation.
OK Symbol and Repair
Symbol are flashing
during stand-by.
OK Symbol and Repair
Symbol are lighting
continuously during
operation.
Repair Symbol is
flashing during
stand-by.
Repair Symbol is
lighting continuously
during operation.
Cause
Acoustic
Battery is low. Battery
Signal tone in
can only supply a limited
every hour
number of shocks.
Battery is low. It can only
"Battery low."
supply a limited number
of shocks.
Signal tone in
Battery is empty.
every hour
"Battery low."
or "Device is not
Battery is empty.
ready for use."
No acoustic
The electrodes are not
message
connected to the device.
The electrodes are not
connected to the device.
"Plug electrodes
into socket."
Wrong electrodes are
being used.
Signal tone in
There is a malfunction
every hour
with the device.
No acoustic
There is a malfunction
message
with the device.

7. Troubleshooting

Action
CardiAid can be used only in
emergencies. Contact Cardia
International or an authorized
service provider immediately
for battery replacement.
CardiAid can be used only in
emergencies. Contact Cardia
International or an authorized
service provider for battery
replacement.
Device cannot be used. Contact
Cardia International or an
authorized service provider
immediately.
Device cannot be used. Contact
Cardia International or an
authorized service provider
immediately.
Plug the electrodes to the
device.
Plug the electrodes to the
device.
Be sure that original CardiAid
electrodes are being used.
Device cannot be used. Contact
Cardia International or an
authorized service provider
immediately.
Device cannot be used. Contact
Cardia International or an
authorized service provider
immediately.