6. Function Test

CardiAid performs an extensive self-test periodically on stand-by status
(cover closed). Additionally, a self-test is performed at the start of each
operation (when cover is opened). The result of the self-test is indicated with
status ındicators: battery symbol, repair symbol and OK symbol. In addition
to these periodical self-tests, functions of CardiAid should be checked on a
regular basis.

6.1 Periods of User Tests

Daily
Check the status indicator symbols of CardiAid daily. If the green OK symbol is flashing in stand-by
position, CardiAid is ready for use. If the red "Battery Symbol" or "Repair Symbol" is flashing, see
Section 7. Troubleshooting. If the problem persists, contact Cardia International or an authorized
service provider immediately.
Semi-annual
Perform a function check every 6 months. (See Section 6.2. Function Check for details)

6.2 Function Check

Perform function check every 6 months as explained below. If you observe an error or discrepancy
in the values, do not use the device and try to solve the problem by yourself by referring to Section
7. "Troubleshooting". If the problem cannot be solved, contact Cardia International or a Cardia
International authorized service provider immediately.
1. Check the status indicator symbols while CardiAid is in stand-by mode (the cover is closed). If the
OK symbol is flashing in stand-by mode, CardiAid is ready for use.
2. Open the cover of CardiAid. If the following conditions are met, the device is ready to use:
• All lights and status indicator symbols light simultaneously for a short period of time.
• Then, OK symbol lights continuously.
• CardiAid starts giving verbal instructions.
3. Close the cover again and make sure that the device is in stand-by mode. If verbal instructions
stop and OK symbol starts flashing, CardiAid is ready for use.
4. Inspect the appearance of the device. Check if the device has external damage. If the device is
damaged, it shouldn't be used.
5. Check whether all accessories are complete and unused. Missing or faulty parts should be
renewed immediately.
6. Check whether the electrode connector is plugged in the socket correctly. If the electrode plug is
not connected correctly, press the plug firmly to the socket.
7. Check whether the electrode plug, electrode cables and the electrodes are in good condition. If
the plug, cables or the pack is damaged, electrodes should be replaced immediately.
8. Make sure that the electrode pack has not passed its expiration date. If it is expired, the
electrode pack should be replaced immediately.