Adjustment With External Weight - KERN and SOHN CXB Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operation
followed by
The sixth adjustment item is displayed, by that way linearisation is completed.
Use
to change to adjustment (see chap. 6.7.2 adjustment CXB-M –models)
After successful linearisation you will have to carry out an adjustment.
In case of an adjustment error or incorrect adjusting weight the display will show an
error message; repeat linearization process.
Cancel linearisation process: press
.

6.7 Adjustment with external weight

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Procedure when adjusting:
In calibrated balances the adjustment function is switch locked. To carry out
adjustment, you will have to throw-over the unlock switch.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing plate.