Checking the device
GB
Zero
Test
On
Off
Zero
Test
MD1X0/200_1a 03/2020
If the result falls outside of the tolerance, the shelf life of the standard
must be checked as well as whether the photometer is in the factory
calibration. If this applies, we recommend returning the device to the
manufacturer or distributor.
Reference standards-Kits for MD 100, MD 110 and MD 200
The reference standards are only used to check the measurement
accuracy for the photometers listed in the test certificate. The calibration
of these devices is stable in the long term under normal working
conditions. The photometer cannot be calibrated with the reference
standard.
The reference standards are delivered in tightly sealed vials.
The target values for each standard are documented in the certificate.
The shelf life of the reference standard is two years from the date of
manufacture, provided it is used and stored correctly. The outside of
the vials must always be cleaned with a clean cloth before use.
Part-
Material
Number
Kit chlorine for devices with tablet/
275650
liquid reagents
Kit chlorine for devices with tablet/
275655
liquid reagents
Kit chlorine for devices with tablet/
275656
liquid reagents
275660
Kit chlorine for devices with tablet /
powder reagents
Kit pH for devices with tablet / liquid
275670
reagents
* Standard value, exact value according to the analysis certificate.
Use of a verification standard kit (see also verification standard
kit, page 44)
press and and hold
press briefly, release
release
range
0,2* + 1,0* mg/l
0,5* + 2,0* mg/l
1,0* + 4,0* mg/l
0,2* + 1,0* mg/l
7,45* pH
45