Télécharger Imprimer la page

Functional Description; Operation; Scroll Memory; Performing Zero - Tintometer Lovibond MD 100 Mode D'emploi

Masquer les pouces Voir aussi pour Lovibond MD 100:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 50

Functional description

GB
On
Off
Mode
Sample
MD1X0/200_1a 03/2020

Operation

Switch the unit on
Selecting a method
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test method, the instru-
ment memorizes the last method used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test method first.
Note:
The description of the methods and reagent handling is not included
in this manual; it is contained in a separate document.
Fluoride method
Caution:
A calibration of the method has to be performed before the first use
4
(see
calibration mode Fluoride). The same batch of SPADNS reagent
solution must be used for adjustment and test. The adjustment process
needs to be performed for each new batch of SPADNS reagent solution.

Performing zero

Fill 24 mm vial with 10 ml sample
Close vial
Place sample vial in the sample chamber.
Pay attention to the positioning.
33

Publicité

loading

Ce manuel est également adapté pour:

Lovibond md 110Lovibond md 200Lovibond md 200 csb